  1. Risk Management Inspection

    0 infractions

  2. Risk Management Inspection

    1 infraction

    • 20. Is reusable equipment disinfected properly?
      • Test strips available onsite at the reprocessing sink are meant for testing pH (inappropriate/incorrect test strips). No other test strips were observed during the inspection to verify high-level disinfectant concentration levels.April 15,2026: Information regarding the appropriate sanitizer test strips was provided during the inspection. The operator indicated that she would place an order and, in the interim, will not reuse the sanitizer solution. Obtain product-specific test strips to verify high-level disinfectant concentration levels. Until the appropriate test strips are obtained and available for use, do not reuse the disinfectant solution, as the concentration cannot be verified to ensure effectiveness.
  3. Monitoring Inspection

    6 infractions

    • 08. Is hand hygiene performed by the personal services worker when required, in a proper manner and are provisions available to perform hand hygiene?
      • Paper towels were not available at the handwashing station or the reprocessing sink area. During the interim period, reusable towels laundered onsite were being used as single-use towels until new supplies arrive.Ensure that single-use paper towels are provided at all handwashing and reprocessing sinks at all times.
    • 15. Is reusable equipment handled and stored properly (Critical)?
      • Reusable semi-critical equipment and drill bits were observed placed on a reusable towel at the nail workstation and on a salon trolley in the spa room. No system or visual indicators were in place to clearly identify whether the items had been reprocessed (clean) or were contaminated/used.Ensure that clean and contaminated equipment are stored separately and clearly identified at all times to prevent cross-contamination.
    • 20. Is reusable equipment disinfected properly?
      • The most recent high-level disinfectant logs available onsite for recording verification of minimum effective concentration for disinfecting semi-critical equipment were dated October, with no year indicated on the records.Ensure that high-level disinfectant logs are maintained accurately and routinely, to demonstrate effective disinfection practices.
    • 20. Is reusable equipment disinfected properly?
      • Test strips available onsite at the reprocessing sink are meant for testing pH (inappropriate/incorrect test strips). No other test strips were observed during the inspection to verify high-level disinfectant concentration levels.Obtain product-specific test strips to verify high-level disinfectant concentration levels. Until the appropriate test strips are obtained and available for use, do not reuse the disinfectant solution, as the concentration cannot be verified to ensure effectiveness.
    • 26. Are cosmetic products appropriate for their intended use and being handled properly?
      • Cosmetic product residue buildup was observed on the pump heads of cosmetic product dispenser bottles in the hair removal services room.Ensure that all cosmetic product dispensers are cleaned regularly and maintained in a clean and sanitary condition at all times to prevent contamination.
    • 27. Is the facility maintained in a visibly clean state and are environmental surfaces cleaned and disinfected properly?
      • Dust buildup was observed on work surfaces at the manicure workstation, and on the salon-trolleys located in the hair removal services room and the lash services room.Ensure that all work surfaces and equipment are cleaned regularly and maintained in a clean and sanitary condition at all times during operational hours.
  4. Monitoring Inspection

    1 infraction

    • 25. Are the equipment manufacturer's instructions for use of specific equipment available and followed?
      • -Manufacturer's instructions for laser machine was not kept in accessible place.Action taken/required:-Staff was instructed to keep the manufacturer's instruction nearby to the machine for staff review if needed.
  5. Monitoring Inspection