Allure Aesthetics Inc.

3222 - 4310 104 Avenue NE Calgary AB T3N 1W3 · Personal Services

3 inspections

  1. Risk Management Inspection

    2 infractions

    • 20. Is reusable equipment disinfected properly?
      • The high-level disinfectant (Revital - OX Resert) was being reused for a period of one week without testing and documentation. - Conduct testing of the high-level disinfectant prior to each use to ensure the minimum effective concentration is achieved. Document all testing results.
    • 25. Are the equipment manufacturer's instructions for use of specific equipment available and followed?
      • A mirror and a portrait with reflective properties were present in the laser treatment room.Remove mirrors and any other reflective surfaces from the laser room to prevent unintended laser beam reflection and ensure operator and client safety.
  2. Risk Management Inspection

    0 infractions

  3. Monitoring Inspection

    5 infractions

    • 20. Is reusable equipment disinfected properly?
      • The high-level disinfectant (Revital - OX Resert) was being reused for a period of one week without testing and documentation. - Conduct testing of the high-level disinfectant prior to each use to ensure the minimum effective concentration is achieved. Document all testing results.
    • 21. Is equipment that is intended to be sterile maintained sterile at the point of use?
      • A pack of expired needles was observed in the storage cabinet.Remove and properly dispose of all expired needles. Ensure that only sterile, unexpired needles are stored and used for client services and implement a system to routinely check and document expiry dates.
    • 21. Is equipment that is intended to be sterile maintained sterile at the point of use?
      • Sterility assurance documentation was not being documented with corresponding client record.- Maintain sterility assurance documentation for all pre-purchased sterile items with the corresponding client record to ensure traceability
    • 25. Are the equipment manufacturer's instructions for use of specific equipment available and followed?
      • A mirror and a portrait with reflective properties were present in the laser treatment room.Remove mirrors and any other reflective surfaces from the laser room to prevent unintended laser beam reflection and ensure operator and client safety.
    • 32. Are general nuisance conditions, other than those noted in other violations, being prevented at this personal services facility (Critical)?
      • The product “MCCM Exosome PDRN” was observed in the facility, the operator informed the inspector that the product is administered via microneedling. The manufacturer’s information lists milk extract without specifying its origin, and the packaging includes the claim “Exosome PDRN is a cellular bioremodeling agent widely used to stimulate cell repair and regeneration of the skin layers.” Under Health Canada’s guidance, products making such claims do not qualify as cosmetics. The product also does not display a Drug Identification Number (DIN) or Natural Product Number (NPN)and there has not been evaluated for safety. Discontinue use of MCCM Exosome PDRN immediately until manufacturer obtains proper Health Canada authorization (DIN or NPN).