Bad Habit Tattoo

2023 99 Street NW Edmonton AB T6N 1M1 · Personal Services

6 inspections

  1. Risk Management Inspection

    0 infractions

  2. Risk Management Inspection

    1 infraction

    • 22. If the facility sterilizes equipment, are all proper sterilization, verification, and documentation processes in place (Critical)?
      • *** Type 5 chemical indicators have expired since 2023-01-06.Corrective Action:Please acquire new type 5 chemical indicators that are not expired.
  3. Monitoring Inspection

    2 infractions

    • 22. If the facility sterilizes equipment, are all proper sterilization, verification, and documentation processes in place (Critical)?
      • *** Type 5 chemical indicators have expired since 2023-01-06.Corrective Action:Please acquire new type 5 chemical indicators that are not expired.
    • 23. If the facility sterilizes equipment, are all proper sterilization, verification, and documentation processes in place (Non-Critical)?
      • *** Sterilized packages are not labeled with load/cycle numbers.Corrective action:Going forward, please ensure every sterilized pouch is assigned and labeled with the load/cycle number. Each package must be able to be tracked to the specific sterilizer and cycle.
  4. Risk Management Inspection

    0 infractions

  5. Demand Inspection

    1 infraction

    • 22. If the facility sterilizes equipment, are all proper sterilization, verification, and documentation processes in place (Critical)?
      • - Biological indicators (BIs) were expired since August 2022.- Physical parameters were not accurately documented sterilization time and temperature of 45 mins and 250F were documented, however the correct parameters for the cycles used 270F for 10 minutes.Corrective Actions:- Acquire and process new BIs.- Correctly document the sterilization temperature and the amount of time the sterilizer maintains that temperature.Recommended Actions:- Use AHS sterilization log for documentation of cycle parameters, since it allow for documentation of external, internal, type 5 chemical indicators, as well as sterilization temperature and time, instead of cycle time.
  6. Initial Inspection

    0 infractions