Incantations Electric Tattoos & Apothecary

10544D 82 Avenue NW Edmonton AB T6E 2A4 · Personal Services

2 inspections

  1. Risk Management Inspection

    0 infractions

  2. Monitoring Inspection

    1 infraction

    • 22. If the facility sterilizes equipment, are all proper sterilization, verification, and documentation processes in place (Critical)?
      • * Physical parameters for each cycle were not being accurately documented. Sterilization times and exposure times do not match the packaged and unpackaged cycles from each of the sterilizers.* External and internal chemical indicator on packages were not documented in the sterilization log.* Biological indicator results were not documented in the sterilization log.Please ensure:- Physical parameters are accurately documented. (Statim 2000 - unpackaged, 3.5mins at 135C / Ritter M9D - unpackaged, 3 mins at 270F, packaged 5 min at 270F)- Verify that all packages have their internal and external chemical indicators have passed and document those results in the sterilization log.- BI results are documented in the sterilization log. Test Vial, pass = no growth, fail = growthControl Vial, pass = growth, fail = no growth