Dr. Bishop & Associates Beautify

230 - 200 Hartell Way SE Calgary AB T2X 4S9 · Personal Services

4 inspections

  1. Monitoring Inspection

    1 infraction

    • 26. Are cosmetic products appropriate for their intended use and being handled properly?
      • One bottle of Alumier AHA Duo Topical liquid-chemical peel was expired-expiry date 12/2025Requirement: Discard this expired chemical peel.
  2. Risk Management Inspection

    0 infractions

  3. Demand Inspection

    3 infractions

    • 12. Are personal services workers at the facility maintaining client records and where applicable, sterility documentation corresponding to that client?
      • The lot number of pre-sterilized items have not been documented. Please record the lot number of pre-sterilized single-use applicators along with the corresponding client information.
    • 20. Is reusable equipment disinfected properly?
      • The isopropyl alcohol currently used for disinfecting the reusable tools and handpieces of the laser equipment does not have a Drug Identification Number (DIN). Please obtain a low- or intermediate-level disinfectant to disinfect the handpieces (points of contact) of the laser devices. Additionally, all reusable tools classified as semi-critical must be disinfected using a high-level disinfectant as a minimum standard. Please ensure that all disinfectants used are approved and have an appropriate Drug Identification Number (DIN).
    • 26. Are cosmetic products appropriate for their intended use and being handled properly?
      • The ultrasound gel bottle is refilled by topping up as needed. Prior to each refill, please wash, rinse, and thoroughly dry the refillable bottles.
  4. Risk Management Inspection

    2 infractions

    • 12. Are personal services workers at the facility maintaining client records and where applicable, sterility documentation corresponding to that client?
      • The lot number of pre-sterilized items have not been documented. Please record the lot number of pre-sterilized single-use applicators along with the corresponding client information.
    • 26. Are cosmetic products appropriate for their intended use and being handled properly?
      • The ultrasound gel bottle is refilled by topping up as needed. Prior to each refill, please wash, rinse, and thoroughly dry the refillable bottles.