Eleve Rx
3517 46 Avenue Beaumont AB T4X 2W3 · Personal Services
2 inspections
- Risk Management Inspection
0 infractions
- Initial Inspection
3 infractions
- 20. Is reusable equipment disinfected properly?
- **There was no disinfectant on site for disinfecting the reusable hydro fascial machine tips.ACTION REQUIRED:1) Purchase a low-level disinfectant for disinfecting the hydro fascial tips between clients.
- 21. Is equipment that is intended to be sterile maintained sterile at the point of use?
- **The microneedling needle cartridges were missing sterility assurance information on the packages. The needle packages were in a zip log bag. The operator stated that they purchased them from the institute that trained them. Cartridge packages were missing manufacturer information, method of sterilization, process indicator etc. ACTION REQUIRED:1) The operator was directed to put the needle cartridges on hold until they are able to provide sterilization assurance documentation with lot numbers that matches those on the cartridges.
- 29. Is the facility properly constructed and maintained to allow for the safe and sanitary provision of personal services (Non-Critical)?
- An uncleanable foam pad observed on the service bed. The operator was directed to remove the foam pad and not to use it because it cannot be cleaned and disinfected.
- 20. Is reusable equipment disinfected properly?