  1. Risk Management Inspection

    0 infractions

  2. Risk Management Inspection

    0 infractions

  3. Initial Inspection

    1 infraction

    • 20. Is reusable equipment disinfected properly?
      • The high-level disinfectant was unavailable for review at the time of the inspection. Please ensure semi-critical equipment is reprocessed with a high-level disinfectant and ensure records are maintained. High level disinfectants must have:- MDL (Medical Device License) -Labelled with the words: sporicidal or high-level disinfectant and Tuberculocidal (TB) or Mycobacterium.Before reprocessing with a high-level disinfectant:-Verify concentration with a test strip.-Maintain records of test date, results, date disinfectant was changed, initials of the worker who performed activity. Steps for reprocessing semi-critical equipment with a high-level disinfectant:-Clean equipment to remove soil.-Rinse to dislodge soil and detergent.- Air dry.-Inspect for cleanliness.-Immerse equipment in high-level disinfectant, ensure no bubble as present. -Follow manufacturer's instructions for contact time, reuse periods, and rinsing requirements. -Store the equipment in a manner that prevents potential contamination. A high-level disinfectant log template, information on disinfectants, and a copy of the Personal Services Standard will be sent via email with the report.